What is a Mass Flow Controller?
Principle and sturacture of mass flow controllers.
What is a Mass Flow Controller?
A mass flow controller is a device used to control the flow rate by measuring the mass flow rate of fluid and gases.
It is possible to control the flow rate stably with high accuracy because it is not required to be corrected due to the
change of temperature or used pressure.
Flow control by volume flow such as flow meters and needle valves, are required to be corrected due to the
change of temperature or pressure when they require a certain degree of the accurate flow rate.
Mass flow controllers and mass flow meters have been used for the field of semiconductor and flat panel display
that require the high accuracy of measurement and control. In addition, they can be used in the wide variety
of processes such as in the field of analysis, fuel cell, liquid crystal, organic EL, biotechnology, food,
environmental measurement, combustion gas control, factory equipment, laboratory equipment and etc.
We aim to design and manufacture products that can be used in a wide range of fields. Therefore, FCON's mass
flow controllers / mass flow meters are designed with improving performance and reliability through
original technology pursuing basic performance.
Principle and Structure of Mass Flow Controller
The mass flow controllers are structured from flow sensor, bypass, valve, and the control circuit as shown in the figure below.
In general, the flow sensor used is called thermal flow sensor. Two resistors generate heat when electric current
flows through these that are wounded on the upstream side and the downstream side, around the stainless
capillary tube.
When no gas is flowing in the capillary tube, the heat of the upstream and downstream sides of the resistor is
kept equilibrium and the flow rate output signal shows zero.
Gas entering from the entrance is divided into the sensor and a bypass. When the gas begins to flow into the
sensor, a temperature difference occurs between the upstream and the downstream resistors, thermal balance is
lost and the temperature distribution of the sensor changes. This change is captured by the bridge circuit as flow
rate output signal. Compare the external flow rate setting signal (0 to 5V) and the flow rate output signal of the
sensor, the flow control valve performs PID operation (proportional, integral, derivative) to match the signal level.
Since the valve opening is fine-tunes and adjust automatically, it is possible to control the flow rate under the
setting condition.
|
|
■Structure of Mass Flow Controller |
■Temperature distribution of Flow Sensor |
Conversion factor
Conversion factor is a number which represents the change in flow rate displayed by the type of gas. Generally, (N2) is a reference (1). |
e.g. Flow O2 to the calibrated Mass Flow Controller with He. How much the actual flow rate of O2 when it is displayed with 100ccm?
Actual flow rate of O2 = a number of display × (C.F of gas/C.F of mass flow controller) = 100 × (0.99/1.40) = 70.71ccm |
● The public standard CF should be used for reference only. |
Gas type |
C.F |
Gas type |
C.F |
N2 (Nitrogen) |
1.00 |
Air |
1.00 |
He (helium) |
1.40 |
CO (Carbon monoxide) |
1.00 |
H2 (Hydrogen) |
1.00 |
CO2 (Carbon dioxide) |
0.74 |
O2 (Oxygen) |
0.99 |
NO (Nitric oxide) |
0.99 |
Ar (Argon) |
1.40 |
C2H2 (Acetylene) |
0.70 |
● Mass Flow Controller is basically adjusted the gas type and flow rate that specify by customers. In case using the other gases, Calculated value by C.F and the actual flow rate may not match. In addition, it may cause abnormal operation or abnormal flow by the differences in the physical properties values of the gas.
● It has also been selling multi-gas multi-range (MG / MR) mass flow controller in response to several kinds of gas and flow rate.
● If you use other gases, please contact the company you bought directly.
Points to keep in mind when using a mass flow controller
|
● Generally, the mass flow controller will not be able to completely shut off the flow of gas, even if the setting of the zero flow and forced closed. Most of the mass flow manufacturers do not guarantee on out of the flow rate control range basically.
● We recommend providing a valve in front of the mass flow controller. By so doing, mass flow controller will set the flow, and open or close the valve. You can completely shut off if you do not shed gas. In addition, the flow rate is not changed on mass flow controller. The same set flow rate will be reproduced.
● Even if the flow rate is changed frequently during passing gas, we recommend providing a valve when you shut off gas completely.
■ Left, valve arrangement. (e.g.) |
To those who are new to Mass Flow Controllers
■ Flow rate setting of the mass flow controller is easy if you use the control power of the dedicated.
■ First, connect the mass flow controller body in the piping and hoses.
■ Adjust in required differential pressure (secondary side air release time: 50 - 300kPa) by the pressure regulator.
■ Connect the control power supply and the mass flow controller in the signal cable.
■ Then, turn on the power to connect the power cable of the control power supply.
■ FCON's control power supply [PA01S or PA01PS] : Set the front switch to SET, turn the volume and set the set flow rate.
■ When flowing gas, the set flow rate is automatically controlled.
● To change the flow rate, turn the volume of the control power supply and adjust the flow rate.
Digital Mass Flow Controller [2000 Series]
|
● Built-in display setting unit, setting control is possible only with the main unit.
● Driving power is DC24V single power supply (At least 500mA AC adapter can also be used).
● Compact size to realize small size and space saving.
● Easy installation to various equipment.
● Digital communication: RS485 communication.
● Analog signal: 0-5V and 4-20mA can be switched.
● Equipped with an abnormal flow rate alarm function.
● Equipped with accumulation function.
● It can be used in a wide range of fields including various facility equipment and experimental research. |
Go to Digital Mass Flow Controller[2000 Series]page |
Mass Flow Controller [1000 Series]
|
● Basic functions in pursuit of the unique technology.
● Flow range: F.S. 10ccm - 100LM
● Compact size.
● The suppression of Flow Surge at the start. (Requires more than 30 seconds of interval)
● Input & Output Signal : Analog (0-5VDC)
● It can be used in a wide range of fields including various facility equipment and experimental research. |
Go to Mass Flow Controller[1000 Series]page |